S E C T I O N 1 1 • T e s t i n g
Proper Calibration of Recirculating Solution Temperature
Confirm proper calibration of recirculating solution temperature by
performing the following steps:
1
Plug the DSTA-40 into the auxiliary outlet located on the
bottom of the Fluid Warmer.
2
Install the DSTA-40 in the heat exchanger position (#1 Block
and #2 Block).
3
Install a test filter in the #4 Block.
4
Turn the Fluid Warmer ON. Allow to warm up until the
DSTA-40 temperature display stabilizes.
5
Compare the DSTA-40 temperature display with the
temperature display on the Fluid Warmer’s Display Panel.
Note:
The Fluid Warmer’s temperature display must read within
0.3°C of the Level 1
®
DSTA-40 TEMPCHECK display on a properly
calibrated unit.
If the DSTA-40 reads 41.7°C and the Fluid Warmer’s
temperature display is within the range of 41.4°C-42.0°C, the
calibration is OK.