78
Version 2.0, Revision N
Cleaning, Disinfection, Maintenance
9
2. Detach the metal ring
in the direction of the arrow
and carefully pull the lamp out of the holder with your right hand
1
2
3
Fig. 64
Installation of the new lamp
1. Carefully slide the metal ring
over the lamp from the left
1
Fig. 65
2. Push the lamp into the holder on the left side until it engages.
3. Screw the metal ring onto the holder, then hold the lamp with both hands and carefully push it
into the clips
4. Reinsert the sliding window.
5. Connect the instrument to the power supply again and switch it on.
Warning
•
If any metallic mercury is released, handle it carefully and dispose of it properly.
•
Dispose of the UVC lamp separately.
9.3.4 Replacing the LED illumination
The LED illumination is designed for maximum life span. In case of damage, please contact Leica
customer service to arrange replacement. For further details, see