Leica CM1860/CM1860 UV
Operating the Instrument
6
3
Status: Short disinfection aborted
The LED for short disinfection flashes, and the UVC lamp is off. The operator must press the
UVC
button
to confirm. The disinfection system goes into ready mode, or, if the lifetime of the UVC lamp has expired,
both LEDs flash alternately.
4
Status: Long disinfection aborted
The LED for long disinfection flashes, and the UVC lamp is off. The operator must press the
UVC
button
to confirm. The disinfection system goes into ready mode, or, if the lifetime of the UVC lamp has expired,
both LEDs flash alternately.
Note
The disinfection mode is aborted as soon as the sliding window of the cryostat is open or the operator
presses the
UVC
button. In the case of a power failure, when the instrument restarts it checks whether
it was running a disinfection cycle, and if so which one. The LED of the determined disinfection cycle
(short or long disinfection) flashes.
6.2
Configuring the desired values
6.2.1 Setting the time
Fig. 15
1
2
Use the function button labeled with a clock symbol to set the time of day.
•
Set the current time using the
Plus
and
Minus
buttons.
•
Pressing and holding the
Plus
or
Minus
button continuously increases or decreases the time
(auto-repeat function).