51
Leica CM1860/CM1860 UV
Daily Use of the Instrument
7
Warning
The safety gloves included in the
P. 26 – 4.3 Standard delivery - packing list)
disposing of the blade!
Lateral movement
If the sectioning results are not satisfactory, the blade holder (on a base here
shifted sideways in order to use another part of the blade, and to benefit from the entire length of the
blade.
To do so, follow these steps:
1. Turn the clamping lever
to the rear (counterclockwise) to release, and then shift
the blade holder sideways to the desired position.
2. To tighten, turn the clamping lever
to the front (clockwise).
Fig. 38
14
Adjusting the anti-roll guide system
You can adjust the height of the anti-roll guide system using the knurled nut
•
If you turn the nut counterclockwise, the anti-roll guide system moves toward the blade.
•
If you turn the nut clockwise, the anti-roll guide system moves away from the blade.
If the anti-roll guide system is in the wrong position relative to the blade, the following problems will
result:
»
The section rolls over the glass insert of the anti-roll guide system
.
L
Error: Glass insert not high enough.
Remedy: Turn the knurled nut counterclockwise until the section is pushed between the blade and
anti-roll guide as shown in
.