72
Version 2.0, Revision N
Cleaning, Disinfection, Maintenance
9
10. Dispose of liquid according to the waste disposal regulations.
11. Allow the cryochamber to dry thoroughly.
12. Insert the disinfected accessories and tools back into the cryochamber only once they are
fully dry.
13. Switch on the instrument.
14. Once a suitable cryochamber temperature has been reached, return the specimens to the
cryochamber.
1
2
3
Fig. 55
Note
collects the condensate that accumulates during defrosting.
Therefore, check the level regularly and empty the container according to applicable laboratory
regulations.
9.3 Maintenance
9.3.1 General maintenance instructions
To ensure a smooth operation of the instrument over a long period of time we recommend the following:
• At least once a year, have the instrument inspected by a qualified service engineer authorized by Leica.
•
Enter into a service contract at the end of the warranty period. For further information, please contact
your local Leica service center.
•
Clean the instrument
DAILY
.