52
Version 2.0, Revision N
Daily Use of the Instrument
7
Fig. 39
8
»
Sections are compressed and block hits the glass insert
after sectioning.
L
Error: Anti-roll guide system is set too high.
Remedy: Turn the knurled nut clockwise until the section is pushed between the blade and anti-roll
1
2
3
Fig. 40
Note
GENERALLY
we recommend pre-adjusting the anti-roll guide system for high section thicknesses
(e.g. 10 µm). – Start from there and work your way down to the desired section thickness in small
increments, readjusting the anti-roll guide system at each increment using the knurled nut.
Assembling the anti-roll guide system/replacing the anti-roll plate
1. Insert the glass into the interchangeable frame and tighten it evenly using the knurled screws
.
2. Insert the shaft
of the metal frame for exchangeable glass inserts from above
into the hole of the swinging arm in such a way that the pin
rests in the notch.
3. Push the white plastic plate
from below onto the shaft