Proline heating thermostat
17 / 123
The bath level is acquired by the SelfCheck Assistant in 8 stages. A permanent display is provided only with the Com-
mand remote control. If the minimum level is undercut, the pump and heater are switched off. The reaction of the
thermostat in case of overfill can be set to simply display a warning, to display a warning and switch off the heater or to
switch off the unit completely with pump and heater.
When the level is too low, with overtemperature, or with other alarms the SelfCheck Assistant switches the heater off
on all poles. The pump is also switched off.
This switch-off under fault conditions is retained, i.e. after the fault is rectified, the fault must be reset (released) on the
Master operating panel with the key.
Other unit functions are described in the appropriate sections and in Section 7. (Starting up).
4.6
Programmer and ramp function
Master Version:
No programmer provided.
Command remote control:
The units are equipped with a programmer function, which enables five temperature/ time programs to be saved. Each
program consists of a number of temperature/ time segments. These also include details of how often the program is to
be executed. Up to 150 segments can be distributed amongst the five programs.
With the ramp function, a rate of change can be directly entered in °C per unit time. (
4.7
Interfaces
Master Version:
The Master unit is equipped with the following sockets at the back of the control head:
For the connection of an external Pt100 temperature sensor (10S).
Two sockets (70S) for the connection of components via the LAUDA equipment bus (cooling section, Command
remote control, external solenoid valve, etc.).
Command remote control:
The Command remote control is equipped as standard with the following sockets:
For the connection of an external Pt100 temperature probe (10S).
Two sockets (70S) for the connection of components via the LAUDA equipment bus (cooling section, Command
remote control, external solenoid valve, etc.)
One RS 232/485 interface (65S) at the back of the Command remote control.
Содержание PROLINE P 5 C
Страница 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Страница 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Страница 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Страница 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Страница 118: ......
Страница 119: ......