30 / 123
Proline heating thermostat
The following always applies: After termination of the relevant settings, they
are accepted automatically after approx. 4 s or
the setting is accepted immediately with the Enter key.
and
Fault signal: Flashing red Alarm LED and acoustic signal.
An acoustic signal can only sound when it has not been intentionally deac-
tivated! (
The bath control occurs via the external temperature probe when the green
LED is lit.
Heating is active when the yellow LED is lit.
Cooling is active. When the setpoint temperature is lowered, it makes take up
to one minute before the blue LED is lit.
The temperature of the external probe is displayed.
Command
Enter key ("Confirm selection") and go back one
level.
End
Soft key function to confirm a selection or input
and to return to the main display window.
Escape key to quit a window without changes
and to go back one level.
Cursor keys for Up, Down, Left and Right.
+
Standby activation (pump and heater are deac-
tivated when the yellow LED is lit).
However, the Timer goes on! See safety infor-
mation on (
Duo key:
Top: Decimal-point key.
Bottom: Key for arithmetical sign.
Soft keys: 5 duo-keys which each have the
function shown in display above them. Soft-key
entries are shown framed in the operating in-
structions. Example: You would like to change
the setpoint temperature, then press the duo-
key under T
set
.
Содержание PROLINE P 5 C
Страница 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Страница 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Страница 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Страница 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Страница 118: ......
Страница 119: ......