Proline heating thermostat
57 / 123
7.8.10
Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
dEF E
3x
to
MNEnu
2x
to
CAL..
3x
Call the offset works setting for the external temperature probe
dEF
E
.
Continue as described in (
7.8.8) for the internal temperature probe.
Command
Default
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to:
Settings
Calibration
extern Pt100
Default
.
The adjacent window appears.
Confirm selection with .
Continue as described in (
7.8.8) for the in-
ternal temperature probe.
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default