84 / 123
Proline heating thermostat
7.14.4
High-level warning or alarm
3 Sec.
Acoustic warning signal sounds for 3 seconds when the liquid level rises so far
that the uppermost switching point of the level sensor has been reached.
Or in case the warning function as described in (
7.14.3) was chosen:
The acoustic signal with dual-tone sounds.
Warning
WXArn103
(high level) appears when the bath contains too
much liquid.
The
WXArn
flashes by turns with the numeral.
In case the alarm function as described in 7.14.3 was chosen:
The acoustic signal with dual-tone sounds.
Red LED
above the fault triangle flashes.
Heater switches off on both poles,
Pump unit is switched off electronically.
Find the cause of the fault. Possible causes may be:
1.
Volume expansion on heating.
2.
Feed to an external vessel may be interrupted so that only return suction
is possible.
3.
Heat transfer liquid taking up moisture.
At Alarm: Press Enter key. Warnings disappear automatically when the cause
is gone.
Also; press this key if the unit has been switched off in the fault state. Warn-
ings disappear automatically when the cause is gone.
Command
High-level warning/ alarm
The display shows
Warning. To release press Enter key
Security 3 Level too high
or
Alarm AL 6: Level too high is shown in the display and signifies that
unlocking is only possible on the Master control panel.
Level warning
WX103
°C
Level alarm
AL 6
°C
Содержание PROLINE P 5 C
Страница 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Страница 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Страница 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Страница 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Страница 118: ......
Страница 119: ......