180
Appendix A: MeasureReady™ FastHall™ Station
MeasureReady™ M91 FastHall™ Measurement Controller
A.6 Maintenance
and
Troubleshooting
This section covers maintenance and troubleshooting. Customer service of the prod-
uct is limited to the information presented in this chapter. Factory trained service
personnel should be consulted if the instrument requires further repair. Damage to
the system can result if these procedures are not done properly. Instructions for con-
tacting Lake Shore and arranging product service are in section 6.8.
A.6.1 Preventive
Maintenance Schedule
Use this table as a foundation in developing a time table for component
maintenance. Tailor the schedule to fit your own system use.
TABLE A-2
Preventive maintenance schedule
Maintenance
Every use
3 months
6 months
12 months
As needed
Maintain a safe, clean work space
×
Clean the sample holder
×
Standard insert middle wiring board
replacement
×
Sample card retainer replacement
×
Prober pin replacement
×
Clean the sample card
×
Clean the electromagnet surface
×
Replace fuses
×
Lubricate o-rings
×
A.6.2 Sample Card
Retainer Replacement
This procedure assumes that the standard insert has been removed from the body,
and the sample card has been removed according to section A.5.3.3. When the sample
card retainer replacement is shipped to you, it is shipped attached to the card
alignment guide. To replace the guide, follow this procedure.
1. Align the card alignment guide (with stem) to the connector on the middle wiring
board, so that the line on the side of the guide aligns with the line on the insert
outer tube.
2. Using a 2.5 mm hex key, screw in the two M3 screws onto either side of the card
alignment guide to secure the guide and stem to the connector on the middle
wiring board.
A.6.3 Sample Card
Maintenance
This section describes maintenance procedures for the sample card.
A.6.3.1 Prober Pin Replacement
Follow these procedures to replace the prober pin on the prober pin sample card. You
will need a number 1 Phillips screwdriver and the 10 mm prober pin sample card
replacement kit.
1. Lay the sample card on the card fixture with the prober pins facing up.
2. Using a number 1 Phillips screwdriver, unscrew the four screws attaching the
prober pins to the card.
3. Remove the four old prober pins.
4. Obtain the new prober pins and appropriate screws from the prober pin replace-
ment kit.
5. Place the prober pin on the card, and screw it in. Do this to all four prober pins.
6.3.1.1 Cleaning Sample Cards
Wear gloves while performing any procedures dealing with the sample cards to avoid
contaminating the sample card or other internal components with oils.
To remove a sample mounted to a sample card with rubber cement, gently push the
side of the sample with a wooden stick or other soft object to shear it from the
adhesive.
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...