113
|
www.lakeshore.com
NOTE:
When the resistivity is omitted or NaN (9.91 e+37),
mobility is not calculated.
Data type is
Number
<blankingTime>
The time, in seconds, to wait for the hardware to settle
before gathering readings.
0.5 to 300 s with a resolution of 0.1 ms
DEFault = 2 ms
MINimum = 0.5 ms
MAXimum = 300 s
Data type is
Number
<sampleThickness>
Thickness of the sample in meters.
0 to 10 e-3 m
DEFault = 0 m
Data type is
Number
<MinimumSnr>
The desired signal-to-noise ratio of the measured
resistance, calculated using measurement average/
error of mean 1 - 1000, or INFinity (9.9 e+37).
DEFault = Infinity
Data type is
Number
<samplingTime>
The sampling time, in seconds, which measurements will be
averaged over to get one sample.
MINimum = 0.010 ms
MAXimum = 1 s
DEFault = 1/60 s
Data type is
Number
Example
HAL:HBAR:DC:START Current,10E-3,10E-3,10E-3,1.0,10,10,0.6,1,0.216,100e-3,
1e-3,30
HALl[:DC]:CONTinue
Continues the DC Hall measurement if it’s in a waiting state.
Command Format
HALl[:DC]:CONTinue
HALl[:DC]:RESet
Summary
Resets the measurement to a not run state, canceling any running measurement.
Command Format
HALl[:DC]:RESet
Example
HALl[:DC]:RESET
HALl[:DC]:RESult:JSON:ALL:SUBstring?
Retrieves a substring of the last run DC Hall measurement results, serialized as JSON.
Query Format
HALl[:DC]:RESult:JSON:ALL:SUBstring? <startIndex>,<length>
Parameters
<startIndex>
The substring starting index, zero based.
Data type is
NR1
<length>
The number of characters to return.
Data type is
NR1
Returns
Substring of the JSON serialized DC Hall measurement result.
Data type is
NAMED
Examples
Query HALl:DC::RESult:JSON:ALL:SUBstring? 0, 42
Query response
{"Setup":{"ExcitationType":1,"ExcitationVa
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...