135
|
www.lakeshore.com
RESistivity:RESult[:STANdard]:RAW?
Summary
Retrieves the measured voltage, current, and calculated current reversed resistance
for a given contact configuration and sample.
Query Format
RESistivity:RESult[:STANdard]:RAW? <sampleIndex>,<contactConfiguration>
Parameters
<sampleIndex>
Sample index, zero based
Data type is
NR1
<contactConfiguration> van der Pauw: R2134, R3241, R4312, or R1423
Hall bar: R5614 or R5623
Data type is
ContactConfiguration
Returns
<measuredVoltagePositive>,<measuredCurrentPositive>,
<measuredVoltageNegative>,<measuredCurrentNegative>,
<CurrentReversedResistance>,<inCompliance>,<voltageOverload>,<currentOverload>
<measuredVoltagePositive> Measured voltage with positive excitation, V+
Data type is
NRf
<measuredCurrentPositive> Measured current with positive excitation, I+
Data type is
NRf
<measuredVoltageNegative> Measured voltage with negative excitation, V-
Data type is
NRf
<measuredCurrentNegative> Measured current with negative excitation, I-
Data type is
NRf
<CurrentReversedResistance> Calculated current reversed resistance.
Data type is
NRf
<inCompliance>
The specified voltage compliance or current limit was
reached.
Data type is
bool
<voltageOverload>
Indicates if, during the measurement, the sample was in a
voltage overload condition.
Data type is
bool
<currentOverload>
Indicates if, during the measurement, the sample was in a
current overload condition.
Data type is
bool
Example
s
Van der Pauw query: RESistivity:RESult:RAW? 4,R3241
Van der Pauw result: 7.504676E-05,9.997895e-06,7.441112e-05,-9.99955e-06,
0.03178598,0,0,0
Hall Bar Query: RESistivity:RESult:RAW? 4,R5614
Hall Bar Result: 7.397222E-05,-7.685482e-10,7.364797e-05,3.068916e-09,
-84.49669,1,0,0
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...