FIGURE 3-7
Contact check with optimization
41
|
www.lakeshore.com
3.8.1.2 Manual Contact Check
Manual contact check allows the user to specify all input parameters manually.
Example:
The example below runs a contact check using current excitation. 11 points will be
taken spanning from +10 µA to -10 µA on the 10 µA range, with the following
settings:
D
Measurement range:
set to 100 mV
D
Compliance limit:
set to 10 V
D
Minimum R2:
set to .9999
D
Blanking time:
set to default
CCHeck:STARt:MANual CURRENT,10e-6,-10e-6,10e-6,100e- 3,10,
11,0.9999,DEF
FIGURE 3-8
Manual contact check