43
|
www.lakeshore.com
3.8.2 Step 2:
Resistivity
This measurement determines the resistivity of the sample between two adjacent
points. In this measurement, the sample should be subjected to zero field. View the
results from a SCPI command or view the high-level results from the M91 front panel.
Results returned are resistivity, F-value, and geometry dependent values.
For hard to measure samples, it is common to increase the number of independent
measurements taken during a single resistivity determination. This permits the stan-
dard error to be calculated and reported. The M91 allows two ways of controlling the
number of samples collected:
D
Specify the number of samples directly.
D
Specify a minimum signal-to-noise ratio (SNR). Samples are continuously col-
lected until the SNR reaches the specified minimum SNR. A maximum number of
samples is also specified to place an upper bound on the number of samples
attempted.
3.8.2.1 Manual Resistivity
Example with minimum SNR:
The example below runs a resistivity measurement using the following settings:
D
Current excitation:
10 µA on the 10 µA range
D
Excitation measurement range:
set to AUTO
D
Measurement range:
set to 100 mV
D
Compliance limit:
set to 1.5 V
D
100 samples will be taken
D
Blanking time:
set to 2.4 ms
D
Sample thickness:
set to 0
D
Minimum SNR:
set to 30
RESistivity:STARt CURRent,10e-6,10e-6,AUTO,100e-3,1.5,100,
2.4,0,30
FIGURE 3-11
Resistivity
Example with specified number of samples:
The example below runs a resistivity measurement using the following settings:
D
Current excitation:
10 µA on the 10 µA range
D
Excitation measurement range:
set to AUTO
D
Measurement range:
set to 100 mV
D
Compliance limit:
set to 1.5 V
D
30 samples will be taken
D
Blanking time:
set to 2.4 ms
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...