127
|
www.lakeshore.com
RESistivity:HBAR:STARt
Summary
Performs a resistivity measurement on a Hall bar sample.
Command Format
RESistivity:HBAR:STARt <excitationType>,<excitationValue>,<excitationRange>,
<excitationMeasurementRange>,<measurementRange>,<complianceLimit>,<width>,
<separation>,<maxNumberOfSamples>, <blankingTime>,<sampleThickness>,
<minimumSnr>,<samplingTime>
Parameters
<excitationType> VOLTage,
CURRent
Data type is
NAMED
<excitationValue>
For voltage excitation -10 to 10 V
For current excitation -100 e3 to 100 e-3 A
Data type is
NRf
<excitationRange>
For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<excitationMeasurementRange> For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<measurementRange>
For voltage excitation, specify the current measurement
range 0 to 100 e-3 A
For current excitation, specify the voltage measurement
range 0 to 10 V
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<complianceLimit>
For voltage excitation, specify the current limit 100 e-9 to
100 e-3 A
For current excitation, specify the voltage compliance 1
to 10 V
Data type is
NRf
<width>
The width of the sample in meters. Greater than 0.
Data type is
NRf
<separation>
The distance between the sample’s arms in meters.
Greater than 0.
Data type is
NRf
<MaxNumberOfSamples>When minimumSnr is omitted or INFinity (9.9 e+37),
the total number of samples to average 1 to 1000.
When minimumSnr is specified, the maximum number of
samples to average 10 to 1000.
DEFault = 100
Data type is
Number
<blankingTime>
The time in seconds to wait for hardware to settle before
gathering readings.
0.5 to 300 s with a resolution of 0.1 ms
DEFault = 2 ms
MINimum = 0.5 ms
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...