<MobilityStandardErrorInMetersSquaredPerVoltSecond>
The standard error of the mobility in meters squared per
volt second.
Data type is
NRf
HALl[:DC]:RUNNing?
Summary
Indicates if the DC Hall measurement is running.
Query Format
HALl[:DC]:RUNNing?
Returns
The measurement is running.
Data type is
bool
HALl[:DC]:WAITing?
Summary
Indicates if the DC Hall measurement is waiting.
Query Format
HALl[:DC]:WAITing?
Returns
The measurement is waiting.
Data type is
bool
HALl[:VDP][:DC]:STARt
Summary
Performs a DC Hall measurement for a van der Pauw sample.
Command Format
HALl[:VDP][:DC]:STARt <excitationType>,<excitationValue>,<excitationRange>,
<excitationMeasurementRange>,<measurementRange>,
<complianceLimit>,<maxNumberOfSamples>,
<userDefinedFieldReadingInTesla>,<withFieldReversal>,
<resistivity>,<blankingTime>,<sampleThickness>,
<MinimumSnr>,<samplingTime>
Parameters
<excitationType> VOLTage,
CURRent
Data type is
ExcitationType
<excitationValue>
For voltage excitation -10 to 10 V
For current excitation -100 e3 to 100 e-3 A
Data type is
NRf
<excitationRange>
For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the range to the best fit range for a given
excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<excitationmeasurementRange> For voltage excitation 0 to 10 V
For current excitation 0 to 100 e-3 A
AUTO sets the measurement range to the best fit range for a
given excitation value.
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<measurementRange>
For voltage excitation, specify the current measurement
range 0 to 100 e-3 A
For current excitation, specify the voltage measurement
range 0 to 10 V
NOTE:
The hardware will be configured to best meet the
desired range.
Data type is
Number
<complianceLimit>
For voltage excitation, specify the current limit 100 e-9 to
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...