44
c
HAPTER
3:
Operation
MeasureReady™ M91 FastHall™ Measurement Controller
D
Sample thickness:
set to 0
D
Minimum SNR:
set to infinity
RESistivity:STARt CURRent,10e-6,10e-6,AUTO,100e-3,1.5,30,
2.4,0,INF
3.8.2.2 Resistivity with Linked Parameters
The resistivity measurement will automatically pull in the parameters listed in con-
tact check. The following parameters will be automatically used as inputs to resistiv-
ity linking: sample thickness. A contact check must be completed prior to running a
resistivity link measurement.
The example below runs a resistivity measurement with linked parameters using the
following optional settings:
D
Measurement range:
set to 100 mV
D
Sample thickness:
set to 0
D
Minimum SNR:
set to 30
D
Max number of samples:
100
RESistivity:STARt:LINK 100e-3,0,30,100
3.8.3 Step 3:
Hall Measurement
To complete a Hall measurement, run a FastHall™ measurement (for van der Pauw
samples only), or the traditional DC Hall method is available for both van der Pauw
and Hall bar samples. See Chapter 1 for more information on Hall theory.
3.8.3.1 FastHall™
The FastHall™ command utilizes the FastHall technology developed by Lake Shore
Cryotronics. It permits a full Hall analysis to be completed without performing an
actual field reversal step, as with a traditional Hall analysis.
3.8.3.1.1 Manual FastHall™
Example:
The example below runs a FastHall measurement using the following settings:
D
Current excitation:
10 µA on the 10 µA range
D
Excitation measurement range:
set to AUTO
D
Measurement range:
set to 10 V
D
Compliance limit:
set to 10 V
D
Max number of samples:
130
D
Field value:
set to 0.5 T
D
Resistivity:
set to 0.1
D
Blanking time:
set to default
D
Number of voltage compensation samples to average:
set to default
D
Sample thickness:
set to default
D
Minimum SNR:
set to default
FASThall:STARt CURRent,10E-3,10E-3,AUTO,10,10,.5,30,.1,DEF,
DEF,DEF,DEF
Содержание Measure Ready M91 FastHall
Страница 6: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 10: ...MeasureReady M91 FastHall Measurement Controller ...
Страница 64: ...54 cHAPTER 3 Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 156: ...146 cHAPTER 4 Computer Interface Operation MeasureReady M91 FastHall Measurement Controller ...
Страница 160: ...150 cHAPTER 5 Options and Accessories MeasureReady M91 FastHall Measurement Controller ...
Страница 174: ...164 cHAPTER 6 Service MeasureReady M91 FastHall Measurement Controller ...