Communications Section
Command
U
A
C
N
Function
Setting user-selected
calibration data or stan-
dard color data (for
memory channel 00)
Sets range of analog
display
Sets memory channel
Inputs ID label
Input format
“U [ x value : y value : Lv value ]”
•
x value or y value can be input in
any of the three formats given in the
examples below for inputting the
value 0.330:
“0.330”
“.330”
“330”
Spaces must not be input before or
after data.
•
Lv value
Acceptable range: 0.01 to 999
Number of characters: Up to 4 (in-
cluding decimal point)
Spaces must not be input before or
after data.
a. In xyLv or T
∆
uvLv display mode:
“A [ range of
∆
x,
∆
y :
range of
∆
Y ]”
b. In analyzer mode, G standard:
“A [ range of
∆
G :
range of R/G, B/G ]”
c. In analyzer mode, R standard:
“A [ range of
∆
R :
range of G/R, B/R ]”
Acceptable range: 0.1 to 99
Number of characters: Up to 3 (in-
cluding decimal point)
Spaces must not be input before or
after data.
“C memory channel number ”
Acceptable range: 00 to 10 (If op-
tional Card inserted, acceptable range
becomes 00 to 99.)
“N memory channel number
[ 10-character ID name ]”
or “N [ 10-character ID name ]”
Acceptable range of memory channel
number: 00 to 10 (If optional Card in-
serted, acceptable range becomes 00
to 99.)
Further information
•
Corresponds to pressing
key in
xyLv or T
∆
uvLv display mode.
•
Cannot be used in analyzer mode.
•
Although
would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing
or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
•
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
•
Corresponds to pressing RANGE
key.
•
Although
would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing
or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
•
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
•
Range data remains in memory even
if CA-100Plus is switched off.
•
Corresponds to pressing MEMORY
CH
or
key.
•
If memory channel number is omit-
ted, memory channel presently be-
ing used will be set.
•
ID label will remain in memory even
if CA-100Plus is switched off.
•
Although
would be pressed to
complete input of data if the opera-
tion were performed using keys,
pressing
or inputting the cor-
responding command “E” is not
necessary. If “E” is input, the
present measured data will be stored
as the target color data.
•
This command will not be accepted
unless “I” has been input and zero
calibration has been completed.
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...