70
[Setting Procedure]
1
5
3
2
4
1
.
Press the MODE key to select the measurement mode
for which you want to set the range.
2
.
Press the
key.
The LCD display section will switch to the menu selection screen.
3
.
Press the
key to open the RANGE setting screen.
Each time the
key is pressed, the screen will switch in the order
PROBE
→
SYNC
→
ID Name input
→
RANGE
→
Measurement
Speed
→
Number of Digits
→
RS232C Baud Rate
→
PROBE.
4
.
Enter the desired range value.
Use the ten-key (
to
,
) to enter the value. (The cursor moves
to the right each time a value is entered.)
Each time the
key is pressed, the cursor moves between x, y and
Lv, between G and B/G, R/G or between R and B/G, R/G.
In this example, the “x, y” range is set to 2.5%, and the “Lv” range is
set to 2.0%.
1
Press the
,
and then
key to set the “x, y” range.
2
Press the
key.
The cursor (_) will move to the “Lv” position.
3
Press the
,
and then
key to set the “Lv” range.
5
.
Press the
key.
The ranges will be set.
* To cancel range setting, press the
key before pressing the
key.
* By default (factory setting), the ranges are set to 10%.
<Error Messages in LCD Display Section>
… For other error messages, refer to page 107.
●
“E4” (after the
key is pressed)
•
Cause : 0.0% was entered.
•
Action : Enter a correct value and then press the
key. The settable range is from 0.1 to 99%.
RANGE x,y Lv
(%) 10 10
_
RANGE G B/G,R/G
(%) 10 10
_
RANGE R B/R,G/R
(%) 10 10
MENU : SELECT
PUSH SPACE KEY
Menu selection screen
Range setting screen
(For xyLv, T
∆
uvLv, u'v'Lv
or XYZ mode)
For analyzer mode (G reference)
For analyzer mode (R reference)
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...