Foreword
Thank you for purchasing the CRT Color Analyzer CA-100 Plus. This instrument is designed for measurement of
color and luminance of various types of color displays including color CRT and PDP displays. Before using this
instrument, please read this manual thoroughly.
Notes on Use
●
This instrument is designed for indoor use only, and should not be used outside.
●
The instrument must never be disassembled as it is composed of precision electric components.
●
Always use the rated power voltage. Connect the AC power cord (for 100-120V or for 200-240V) to an AC
outlet. Make sure that the voltage is within
±
10% of the rated power voltage.
●
This instrument is classified as Pollution Degree 2(equipment which may cause temporary electrical hazards
due to contamination or condensation, or products which are used in such an environment).
●
This instrument is classified as Installation Category II (the specified commercial power voltage should be
used).
●
Do not use the instrument at altitudes of higher than 2000m.
●
The instrument must not be used if foreign matter such as water and metal objects enter it, doing so is very
dangerous.
●
The instrument should not be used in certain environments, such as near a heater which will cause an excessive
rise in its temperature resulting in breakdown. Therefore it should not be used in such an environment. It should
be used in well-ventilated areas, and care should be taken not to allow the vent holes to become blocked.
●
The instrument must not be used in areas subject to rapid changes of temperature, to avoid condensation.
●
The instrument must not be used in areas where there is an excessive amount of dust or where the humidity is
excessively high.
●
The instrument should be used at ambient temperatures of between 10 and 28˚C and humidity of 70 % relative
humidity or less. Be aware that to use it beyond this condition may make it degrade the performance.
●
The instrument must not be exposed to excessive impact and vibrations.
●
The AC power cord must not be pulled or bent excessively nor must excessive force be exerted on it. Doing so
may result in wire breakage.
●
The AC power cord must not be connected to an AC line on which excessive noise is present.
●
The instrument and personal computer must be grounded.
●
If any irregularity or abnormality is found, turn OFF the power immediately, disconnect the AC power cord,
and refer to “Troubleshooting Guide” on page 111.
●
Should the instrument break down, do not try to disassemble by yourself. Contact a Konica Minolta authorized
service facility.
●
Zero Calibration shall be made only after duration of 30 minutes or more passed since the power supply switch
is turned ON if the measuring display has brightness degrees of the following.
●
Measuring Probe
CA-P02, CA-P05
0.5cd/m
2
or less
●
High Luminance Measuring Probe
CA-PH02, CA-PH05
1.0cd/m
2
or less
Notes on Storage
●
The instrument should be stored at temperatures of between 0 and 28˚C (70 % relative humidity or less and no
condensation) or at temperatures of between 28 and 40˚C (40 % relative humidity or less and no condensation).
It is recommended that it be stored with a drying agent (such as silica gel) at approximately room temperature.
To store it in areas subject to high temperatures and high humidity may make it degrade the performance.
●
Take care not to allow condensation to form on the instrument during storage. In addition, pay attention to rapid
temperature changes during transportation to the storage area to prevent condensation.
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...