[Preparation]
1
.
Press the MODE key to select xyLv measurement mode.
2
.
Press the MEMORY CH
and
keys until the memory
channel where you want to perform user calibration ap-
pears.
A memory channel other than CH00 must be selected.
3
.
Place the measuring probe with the display and take mea-
surement.
Set the display so that it can display four colors (RGBW) whose xyLv val-
ues are known.
4
.
Press the
key.
The LCD display section will switch to the user calibration input screen.
5
.
Enter the emission characteristic of R and calibration val-
ues (x, y, Lv).
1
Place the measuring probe with the display, which is now emitting mono-
chrome light of R.
Currently measured values will be displayed.
2
While the probe is placed with the display, press the HOLD key.
The measured values will be hold and the HOLD LED lights up.
3
Press the
key.
The LCD display section will switch to the R calibration value input
screen.
4
Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for
white calibration (see step 7 in “Performing White Calibration” on page
52).
5
Press the
key.
The LCD display section will return to the user calibration input screen,
with the “*” mark displayed on the left of “R”.
6
Press the HOLD key to resume measurement.
6
.
Enter the emission characteristic of G and calibration values (x, y, Lv).
1
Place the measuring probe with the display, which is now emitting monochrome light of G.
Currently measured values will be displayed.
2
While the probe is placed with the display, press the HOLD key.
The measured values will be hold and the HOLD LED lights up.
3
Press
key.
The LCD display section will switch to the G calibration value input
screen.
4
Enter calibration values (x, y, Lv).
Enter them in the same way as when you enter W calibration values for
white calibration (see step 7 in “Performing White Calibration” on page 52).
5
Press the
key.
The LCD display section will return to the user calibration input screen,
with the “*” mark displayed on the left of “G”.
6
Press the HOLD key to resume measurement.
“
*
” mark is displayed.
CH01
U-CAL
P1
W *R
G
B
CH01 x
y
Lv P1
0000 0000
0.00
_
CH01
U-CAL
P1 W *R *G
B
CH01
U-CAL
P1
W
R
G
B
CH01 x
y
Lv P1
0000 0000
0.00
_
CH01
U-CAL
P1
W *R
G
B
“
*
” mark is displayed.
CH01 EXT
Ad P1
[
]
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...