2. Instrument Status at Power-ON
The instrument has been set prior to factory shipment so that it will be set as follows when the POWER switch is
set to ON.
1
Measurement mode
Page 40
xyLv mode
2
Memory channel no.
Page 48
CH00
3
Target color
Page 64
x = 0.3127 y = 0.3293 Lv = 40.00 (cd/m
2
)
4
PROBE
Page 43
P1
5
SYNC mode
Page 36
EXT mode
6
ID name
Page 66
Consists of Blank spaces only.
7
Analog display range value
Page 68
10% (all ranges)
8
Measurement speed
Page 36
FAST
9
Number of display digits
Page 42
3 digits
10
RS232C baud rate
Page 85
9600bps
11
Calibration data (stored) in CH00 to CH99 Page 50
6500K (D65) Konica Minolta’s standard data
12
CA-200 mode/CA-100 compatible mode
Page 15
CA-100 compatible mode
13
Luminance unit
Page 32
cd/m
2
<Changing the Instrument Status at Power-ON>
Change necessary parameters and press the
key for more than five seconds. A bleep will sound, followed by a
whistling sound when the settings are saved. The instrument will start with the new settings when the power is
turned ON next time. (The selected mode and memory channel etc. will be stored in the instrument’s memory, and
they will remain effective even if the POWER switch is set to OFF.)
* For details, refer to the pages given in the above table.
Changing Method for parameters
1
and
2
1
Measurement mode ......... Press the MODE key.
2
Memory channel .............Press the CH
and
keys.
Changing Method for parameter
3
3
Target color value ............The current target color will be changed and then
enter a target color, or select user calibration or
enter the RGB emission characteristic for ana-
lyzer mode.
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...