Settings Section
3
.
Enter target color values (x, y, Lv).
For x and y, a value 10000 times the calibration value must be entered.
Use the ten-key (
to
,
) to enter the value.
The cursor moves to the right each time a value is entered.
Each time the
key is pressed, the cursor moves in the order x
→
y
→
Lv
→
x.
In this example, x=0.3300, y=0.3000 and Lv=39.50 are entered.
1
Press the
,
,
and then
key to enter the “x” value.
2
Press the
key.
The cursor (_) will move to “y”.
3
Press the
,
,
and then
key to enter the “y” value.
4
Press the
key.
The cursor (_) will move to the “Lv” position.
5
Press the
,
,
,
and then
key to enter the “Lv” value.
4
.
Press the
key.
The target color will be set to CH00.
* To cancel target color setting, press the
key before pressing the
key.
* To view the target color you set, press the MR key. (For details, refer to page 73.)
* By default (factory setting), x=0.3127, y=0.3293 and Lv=40.00 (cd/m
2
) are set for the memory channels for which no target color has
been set.
<Notes when Setting/Changing the Target Color>
●
The
key may not be operable if “OVER” is displayed on the LCD display section.
●
Note that the target color is common to all measurement modes (xyLv, T
∆
uvLv, analyzer, u'v'Lv, XYZ).
●
Never press the following keys during target color setting.
Doing so will cancel setting and activate the mode corresponding to the pressed key.
( 0-CAL , MODE , REMOTE , MR , MEMORY CH
,
,
)
CH00 x
y
Lv P1
3300 3000 160.0
_
CH00 x
y
Lv P1
3300 3000 39.50
_
CH00 x
y
Lv P1
3300 3000 39.50
_
CH00 x
y
Lv P1
3300 3293 160.0
_
Содержание CA-100PLUS
Страница 24: ...22...
Страница 46: ...44...
Страница 76: ...74...
Страница 100: ...98...
Страница 108: ...106 Dimension Diagram Main Body Measuring Probe 60 ISO 5mm screw depth 6mm...
Страница 123: ...121 Explanation Section...
Страница 124: ......
Страница 125: ...9222 1891 13 2002 KONICA MINOLTA SENSING INC AEIDPK 10 Printed in Japan...