61
Maintenance and care
BlueShadow Detector 50D Instructions V7659
8.4 Replacing the flow cell
UV light will cause the flow cells to become blind with time (solarization),
making them no longer suitable for use .
The device has been switched off .
The power plug has been pulled .
The capillaries have been removed .
Process
Figure
1.
Remove capillaries at the inlet
1
and outlet
2
of the flow
cell .
2.
Unscrew the knurled-head
screws
3
of the cover plate
4
. Hold the flow cell securely
with your hand during this
procedure .
3.
Pull out the slide with the flow
cell
5
towards the front .
4.
Remove the flow cell upward .
5.
Insert new flow cell from ab-
ove into the flow cell holder .
Keep holding the flow cell .
6.
Push the flow cell holder
against the detector again .
7.
Screw the knurled-head
screws back into the cover
plate
4
and tighten .
Fig. 18
Replacing the flow cell
The flow cell is assembled .
Connect the capillaries .
Prerequisite
1
2
3
4
5
3
Result
Next step