60
Maintenance and care
BlueShadow Detector 50D Instructions V7659
Process
1.
Using the allen screwdriver, loosen the threaded ring
1
.
2.
Remove the cover
2
and the spacer
3
(not part of all flow cells) .
3.
Using tweezers, pull out the pressure bushing
4
and the light-
guide
5
.
4.
Carefully push the light guide out of the holder and strip off the
seal . The seal ring needs to be renewed during every cleaning of
the fiber optics .
5.
Clean the light guide with a clean, soft cloth or with an appropriate
solvent in an ultrasonic bath . Make sure that the clean light guide
is not touched by fingers .
6.
Assemble the flow cell .
7.
Using the allen screwdriver, tighten the threaded ring
1
.
The flow cell lens is clean .
Note:
If the cleaning of the light guide does not have the desired effect,
the light guide has to be replaced .
Mount the flow cell onto the detector .
Result
Next steps