28
Installation and initial startup
BlueShadow Detector 50D Instructions V7659
Process
1.
Insert the depressor tool
3
in an upper small opening at the front
of the pin header
1
.
2.
Insert the cable into the opening
2
underneath the inserted de-
pressor tool .
3.
Pull out the depressor tool .
Check whether the cables are tightly attached . Press the pin header onto
the connector . Finish the installation . Bring the device into operation .
4.8 Analog control
Using the analog port, you can control the wavelength by changing the
applied voltage . A Control Unit is required in order to select the option
ANA LOG in the SETUP menu .
To use the analog port for controlling the detector, you have to set a zero
point and enter a scaling value .
Zero point at 0 V = 000 nm
Scaling: 100 nm per Volt
If 5 V voltage is applied, the wavelength is 500 nm .
4.9 Integrator Port
The integrator connector sends measuring signals from the detector .
Non-bipolar
1 channel
0 to 5 V
DAC 20 bit
Scalable
Offset is adjustable
4.10 Switching on
The device is switched on by the power key on the back of the device .
Installation has been completed .
Procedure
Process
Figure
1.
Insert the plug into the male
connector
2
at the rear panel .
2.
Plug the other end into an
electric socket .
3.
Switch on the device at the
power switch
1
.
Fig. 12
Power switch with male con-
nector
Next steps
Example
Prerequisite
1
2