39
Operation
BlueShadow Detector 50D Instructions V7659
5.5 GLP menu
The
GLP (Good Laboratory Practice) menu is for information purposes
only . The submenus provide information about the device history, deliver
an overview of the device configuration, and inform about the condition
of detector and lamps .
Navigate through GLP menu
1.
Select the device and lamps GLP menus .
2.
Use the horizontal keys
to navigate through the submenu .
3.
If no buttons are pressed, the display returns to the main menu after a
short time .
4.
The menu contains no editable parameters .
Men
u
E
xplanation
Figure
DEVICE GLP
Serial number
Detector serial number
Operating time
Detector operating time
Motor Working
Grating motor operating (move-
ment) time
CALIBRATION
INFO
Menu shows the line density of
the optical grating as well as the
fine deviation of the zeroth spec-
tral order . Both values are used
for exact wavelength calibrat-
ion of the optical grating of the
detector .
VALIDATION
INFO
Menu shows the results of the
self-test for optical bandwidth
and stray light rejection level
(maximum order filter absor-
bance) as well as the validation
of spectral positions of optical
absorption peaks of Holmium
oxide and optical intensity peaks
(H-beta spectral line) .
Firmware
Version/Display
firmware
Currently installed Firmware Ver-
sions for detector and detector
display .
Install. Date
Service Date
Display of the initial installation
and last service .
Procedure