
76
97-1129-01 Rev. D
www.kistlermorse.com
5. Gross Units
Over
Calculated gross weight (or selected
engineering unit) exceeds seven active digits
(9999999).
yes
note:
Write down all existing system parameters before
you change
Disp Format
or
Disp Units
. If you change
Disp Format
or
Disp Units
, you must adjust system
parameters to correspond to new
Disp Format
or
Disp
Units
.
1. Check
Disp Format
[11]. ensure format is consistent
with maximum expected value. For example, xxxx.
xx format yields maximum value of 99999.99, while
xxxxxx00 format yields maximum value of 999999900.
2. Check
Disp Units
[13]. ensure units are consistent
with maximum expected value. For example, a number
in tons represents 2000 times more weight than the
same number in pounds; a number in barrels represents
42 times more volume than the same number in gallons.
3. Check calibration. Recalibrate if required.
6. net Units Over
Calculated net weight (or selected
engineering unit) exceeds seven active digits
(9999999).
yes
note:
Write down all existing system parameters before
you change
Disp Format
or
Disp Units
. If you change
Disp Format
or
Disp Units
, you must adjust system
parameters to correspond to new
Disp Format
or
Disp
Units
.
1. Check
Disp Format
[11]. ensure format is consistent
with maximum expected value. For example, xxxx.
xx format yields maximum value of 99999.99, while
xxxxxx00 format yields maximum value of 999999900.
2. Check
Disp Units
[13]. ensure units are consistent
with maximum expected value. For example, a number
in tons represents 2000 times more weight than the
same number in pounds; a number in barrels represents
42 times more volume than the same number in
gallons.
3. Check calibration. Recalibrate if required.
7. Grs Under
Range
Calculated gross weight (or selected
engineering unit) is negative and exceeds
four active digits (9999).
yes
1. Check calibration. Recalibrate if required.
2. One or more sensors may be damaged. Follow
troubleshooting procedures detailed in sensor manual
to locate damaged sensor(s). Replace if required.
8. ADC
Ove
Sensor input is causing counts to go above
2,097,151.
yes
1. Check wiring of sensors to junction box and junction
box to SVS2000. Correct loose, damaged (shorted), or
incorrect wiring.
2. One or more sensors may be damaged. Follow
troubleshooting procedures detailed in sensor manual
to locate damaged sensor(s). Replace if required.
3. Check power supply by measuring Ex and
-ex on Power Supply PCB.
• If SvS2000 is wired to km half-bridge sensors with no
IS Barriers, verify voltage is approximately 12 v.
• If SvS2000 is wired to foil-gage sensors, verify voltage
meets sensor manufacturer’s requirements. If voltage is
incorrect, replace Power Supply PCB.
note: If system includes IS Barriers, check power
supply by measuring Sense and -Sense.
4. Measure voltage Sig and –Sig terminals on
Power Supply PCB. This reading is actual signal coming
back from sensor. For K-M sensors, reading should be
less than 1 v:
• If reading is less than 1 v, verify that Sensor gain
[63] is set as described in Chapter 10, Sensor menu.
If Sensor Gain is set correctly, A/D converter may be
damaged. Contact KM.
• If reading is greater than 1 v, recheck wiring, sensors,
and power supply as described in Steps 1 through 3.
Содержание SVS2000
Страница 1: ...SVS2000 Installation Operation Manual IOM97 1129 01 Rev D...
Страница 2: ......
Страница 10: ...6 97 1129 01 Rev D www kistlermorse com...
Страница 11: ...7 www kistlermorse com 97 1129 01 Rev D...
Страница 12: ...8 97 1129 01 Rev D www kistlermorse com...
Страница 13: ...9 www kistlermorse com 97 1129 01 Rev D...
Страница 14: ...10 97 1129 01 Rev D www kistlermorse com...
Страница 20: ...16 97 1129 01 Rev D www kistlermorse com Figure 3 Wiring Diagram Current Output Card venture...