MKIV OPERATOR'S MANUAL
80
ICFJJJYY.NNN - INSTRUMENT CONSTANTS
The operating state file controls many of the operating parameters of the Brewer. The left column
lists the actual value written in the OP_ST.nnn file. The middle column is the BASIC variable name
used in the Brewer software to contain this value, and the right column is a description of the value's
meaning.
#
Example
Name
1
0
Ozone temperature coefficient for slit 1
2
-.3428
Ozone temperature coefficient for slit 2
3
-.6509
Ozone temperature coefficient for slit 3
4
-1.3636
Ozone temperature coefficient for slit 4
5
-2.4043
Ozone temperature coefficient for slit 5
6
.1
Micrometer steps per degree
7
.3407
Ozone on ozone ratio
8
2.35
SO
2
on SO
2
ratio
9
1.1417
Ozone on SO
2
ratio
10
2187
ETC on ozone ratio
11
3340
ETC on SO
2
ratio
12
.000000042 Dead time (seconds)
13
286
Wavelength calibration step number
14
14
Slit mask motor delay
15
1685
Umkehr offset
16
0
Neutral density of filter 0 (divide by 10
4
to get attenuation)
17
4440
Neutral density of filter 1 (divide by 10
4
to get attenuation)
18
10320
Neutral density of filter 2 (divide by 10
4
to get attenuation)
19
14120
Neutral density of filter 3 (divide by 10
4
to get attenuation)
20
21230
Neutral density of filter 4 (divide by 10
4
to get attenuation)
21
25800
Neutral density of filter 5 (divide by 10
4
to get attenuation)
22
2972
Zenith motor steps per revolution
23
mkiv
Brewer model type
24
1
COM port number
25
0
ozone temperature coefficient for mercury exit slit
26
0
NO
2
temperature coefficient for mercury exit slit
27
0
NO
2
temperature coefficient for slit 1
28
0
NO
2
temperature coefficient for slit 2
29
0
NO
2
temperature coefficient for slit 3
30
0
NO
2
temperature coefficient for slit 4
31
0
NO
2
temperature coefficient for slit 5
32
1
ozone micrometer offset
33
0
Not used.
34
242
Ozone filter wheel #3 position
35
-3
NO
2
absorption coefficient
36
765
NO
2
direct sun ETC
37
740
NO
2
zenith sky ETC
38
2511
NO
2
micrometer offset
39
178
NO
2
filter wheel #3 position
40
2510
Ozone/ NO
2
mode change distance for micrometer
41
0
Not used.
42
0
Not used.
43
2669
Micrometer zero position
44
250
number of motor steps to open iris
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...