SPECTROPHOTOMETER UNPACKING AND SETUP
1.
Open the Brewer crate and inspect the contents - at least the following items will be found:
- Brewer Spectrophotometer
- AC Power Cable, BA-W12
- Data Communications Cable, BA-W68
- RS422/RS232 Data Set and AC Power Converter
- Manuals (Operator’s, Service, Final Test Record)
- Basic Spares Kit, BA-C112
- Brewer System Diskettes (three diskettes)
2.
Remove the Brewer Cover by unlatching the four latches and lifting the cover off the Base.
Remove the protective foam from under the Viewing Window, and inspect the Brewer for
loose or broken parts, or disconnected cables.
3.
Connect the AC Power Cable to the appropriate connector as per the markings on the
Brewer Cover, plug the other end into a source of AC power, and press the Power Switch.
4.
Observe that the green Power Lamp illuminates, and that activity occurs as the Brewer
Motors initialize.
5.
Place a few packages of Desiccant (Silica Gel) inside the Brewer and replace the Brewer
Cover.
6.
Turn off the Brewer Power Switch and disconnect the power cable.
3.2
TRIPOD UNPACKING AND SETUP
Refer to Figure 3.6.
1.
Open the Tripod crate and locate the following:
- Tripod Stand
- Tripod tie-down kit (this kit will be used in the final assembly)
2.
Assemble the Stand by spreading the three legs such that the three adjusting pads are
resting flat on the floor.
3.
With the three bolts provided, attach the support bars to the fixings beneath the top flange.
3.3
AZIMUTH TRACKER UNPACKING AND SETUP
Refer to Figure 3.3.
1.
Open the Azimuth Tracker box, remove the Tracker, and inspect it for damage.
2.
Mount the Tracker onto the Tripod and secure it with the bolts provided.
3.
Remove the front and rear covers from the Tracker, and note the spare fuses and mounting
bolts taped to the inside wall of the Tracker.
4.
Locate the Safety Switch, which can be found inside the Tracker housing, mounted on the
diagonal support bar.
The purpose of the Safety Switch is to prevent damage to the Power and Communications
Cables by removing power from the drive motor in the event that the Tracker attempts to
rotate too far.
This Switch is activated by a steel cable and a nylon cord which tighten as the Tracker
rotates.
3. BREWER SYSTEM SETUP
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...