MKIV OPERATOR'S MANUAL
50
3. RS TEST:
RUN / STOP ratios for wavelength 0 and 2-6 should be 1.000 +/- 0.003
If larger deviations are found, then confirm them to be repeatable before taking any
corrective action.
4. DT TEST:
The two summary values should be within a few (5) nanoseconds of the values given in the
Final Test Record.
If larger deviations are found, then confirm them to be repeatable before attempting any
corrective action.
6. RESET
The Reset portion of the printout gives values of Azimuth, Zenith and Micrometers zeroing
discrepancies.
Inspect these discrepancies and confirm that they are repeatable prior to attempting any
corrective action.
7. AP PRINTOUT:
Inspect the monitor values and look for consistent deviations and trends from the Final Test
Record, or from previous values.
8. RL PRINTOUT
The Log file printout should be inspected for any errors that the FirmWare detected, or
thought it detected.
If the errors persist then KIPP & ZONEN should be consulted for advice.
9. OBSERVATION DATA:
Data file summaries should be inspected to ensure that they are present, and data should be
reviewed as per instructions from the Scientific Authority for the Brewer station.
5.2
WEEKLY TASKS
There are a number of tasks which should be performed and results recorded at least weekly.
1. Steps per Revolution (Section 6.14)
2. Solar Siting (Section 7)
5.3
INFREQUENT TASKS
This section deals with true ‘infrequent’ tasks, as well as those that are more for the integrity of the
data than for the operation of the Brewer. The timing of many of these tasks is usually determined
by the “Scientific Authority” of the observation site.
1. UV Stability Check (Section 6)
2. Mechanical Checks
3. UV Calibration
4. Ozone Calibration
5.4
MINOR MAINTENANCE
Routine maintenance deals with periodic checks and adjustments that the Operator should do to
ensure correct operation of the Brewer. These items are not failures, but may lead to failures, or
result in poor data if not attended to.
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...