16-10-200716-10-2007
MKIV OPERATOR’S MANUAL
119
The <name>s
<name>
format of <value> Effective
meaning of <value>
BREWER.ID
integer value (0-
65536)
next warm start
or USECONFIG
This is the Brewer ID used to manage multidrop
protocol. Copied to NVRAM on initialization.
CLOSE.TIME
Float seconds
Immediate
number of seconds before the PMT window
closes when we start to move the slitmask
IMMINENT
float seconds
Immediate
number of seconds before the PMT window
closes that we should start to watch PMT
window closely so we move the motor at the
proper time
LAMP.RESET.TIME
float seconds
Next time lamp is
turned on
The amount of time before the lamp will be
automatically turned off.
LAMP.CONV.CURRENT
[<lamp>]
float amps
Immediate
conversion constant for lamp current as
measured at A/D
LAMP.CONV.VOLTAGE
[<lamp>]
float volts
Immediate
conversion constant for lamp voltage as
measured at A/D
MODEL
unsigned integer
immediate
the 'mark' number of the brewer
MOISTURE.LIMIT
float grams/cubic
meter
on reset
Permitted maximum in watermark recording of
moisture content inside the Brewer.
MOTOR.ADDRESS
[<motorId>]
Unsigned integer (0-
11)
Immediate
Identifies the 751 used to talk to this motor
MOTOR.CLASS [<motorId>]
one of:
NOMOTOR
MICROMOTOR
TRACKERMOTOR
STANDARDMOTOR
SENSORLESSMOTO
R
Immediate
The type of motor at this motorId and hence the
algorithm used to initialize it.
MOTOR.INITIAL[<motorId>]
Signed integer steps Immediate
position of motor from step 0 after initialization
MOTOR.MAX.ACC [<motorId>]
Unsigned integer
paces/(256*tick)
2
on reset
maximum absolute value of motor acceleration
to be permitted
MOTOR.MAX.JERK
[<motorId>]
Unsigned integer
paces/(256*tick)
3
on reset
maximum absolute value of motor acceleration
to be permitted inside a time slice
MOTOR.MAX.SEARCH
[<motorId>]
Signed integer paces on reset
max number of paces before search stopped
MOTOR.MAX.VEL[<motorId>]
Unsigned integer
paces/(256*tick)
on reset
maximum absolute value of motor velocity to be
permitted
MOTOR.MAX.POS [<motorId>] Signed integer paces on reset
maximum absolute value of motor velocity to be
permitted
MOTOR.MIN.POS [<motorId>] Signed integer paces on reset
minimum value of motor position to be
permitted
MOTOR.ORIGIN [<motorId>]
Signed integer paces on reset
position of step 0 after initialization
MOTOR.REF.PLAY [<motorId>] signed integer paces on reset
uncertainty of position of motor with respect to
reference sensor when sensor is detected at
high speed and arbitrary direction.
MOTOR.RESET.POS
[<motorId>]
signed integer paces during motor init
position of motor to be paced at before
commencing motor init to ref sensor
MOTOR.SLOPE [<motorId>]
signed integer
paces/step
Immediate
number of paces to a step (see M command)
MOTOR.SLOW.VEL
[<motorId>]
unsigned integer
paces/(256*tick)
during motor init
value of motor velocity to be permitted during
final approach of search. If equal to max vel init
routine is shortened.
MOTOR.STOP.METHOD
[<motorId>]
int one of
1, 3 none
0 reduced
2 full
on reset
power applied for stopped motor
MOTOR.TIME.OUT
[<motorId>]
float seconds
Immediate
amount of time that a motor is given to complete
movement before timing out.
OPEN.TIME
float seconds
Immediate
number of seconds after PMT window closes
before we open the PMT window
PMT.WINDOW.RESOLUTION float seconds
Immediate
seconds in a window timing tick
APPENDIX H COMPUTER / BREWER INTERFACE (TELETYPE)
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...