MKIV OPERATOR'S MANUAL
40
O3
Change to Ozone Mode
The O3 command changes the operating mode of the Instrument from the NO
2
to the O
3
mode. The change involves rotating Filter Wheel #3 such that the NISO
4
+UG-11 order filter
is moved into the optical path and the BG-12 filter is moved out. In addition, the micrometer is
moved such that the Ozone wavelengths pass through the exit slits. See also N2.
OZSUM Ozone Summary
OZSUM reads the daily Date (B) file and prints out the summary results of the day’s O
3
observations. In addition an entry is made in the OZOAVGYY.nnn file.
The OZSUM command is usually invoked as part of the ED command.
See Appendix A for an example of the OZOAVGYY.nnn file.
PB
Data Playback
Command Syntax: PB BJJJYY.
The PB command allows any previous day’s data to be printed using the current temperature
coefficients.
PD
Print to Disk
PD directs printing to a “DJJJYY.nnn” file on the data drive rather than to the line printer.
See also PN, PF.
PF
Printer Off
PF turns off all printing. Raw data continues to be recorded. See also PN, PD.
PN
Printer ON
PN directs printing to the line printer. See also PD, PF.
PO
Printout Instrument Constants
PO generates a printout of selected instrument constants.
PZ
Point to Zenith
PZ results in the Zenith Prism being pointed to a Zenith angle of 0° (straight upwards).
QS
Quick Scan
QS is used in conjunction with the UV Stability Check Kit, and gives an indication of the
stability of the instrument in the intensity measurement of UV over the range 290nm to
325nm.
When the equipment has been set up as per the instructions in Section 6 of this manual, and
the QS command issued, a report is generated which shows the stability of the instrument at
3.5nm increments using the first generated set of readings as a reference.
For the test, the iris is opened, FW#1 is set to position 3, and FW#2 is set to position 1.
See Appendix A for the LAMP_LLL.nnn file, and Section 6 for more on QS.
RE
Reset
RE initialises all of the Brewer motors (similar to the Power On initialisation), and moves them
to positions as defined in the Instruments Constants File.
Following a RE, the iris is closed, FW#1 is in position 1, FW#2 is in position 3, and FW#3 is
in the position of the appropriate filter.
REP Report
REP displays (or prints) the SLOAVG, HGOAVG, APOAVG, DTOAVG, MIOAVG, and
RSOAVG over a user selected range of days.
If the instrument is in the N2 mode, an equivalent NO
2
report is generated.
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...