16-10-200716-10-2007
MKIV OPERATOR’S MANUAL
57
General Instructions:
1. Three lamps should be scanned bi-weekly. If all three lamps give approximately the same
results, then the test is complete. If one lamp gives results that are different from the other
two, then a fourth lamp should be chosen and the scan run yet another time. The lamp
giving the ‘odd’ reading should be noted and not used again.
2. Once every two months, all five lamps should be scanned to make sure they all give the
same results so availability of spares is assured.
3. In the event that the % difference for all lamps is much greater than 5%, it may be necessary
to perform other scans which will be used to create new response files for the instrument.
See Appendix C.
4. Perform the TU test prior to changing the response file to make sure that the Zenith Prism is
properly aligned. See section 4.
6. UV STABILITY CHECK - QS
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...