16-10-200716-10-2007
MKIV OPERATOR’S MANUAL
61
8
BREWER SCHEDULES - SE, SKC, SK
A most important feature of the Brewer is its capability to run for longs periods of time in an
unattended state.
The Brewer system is able to achieve this Automatic operation the use of Schedules Operation.
A schedule is simply a file containing a list of instructions which the Brewer software executes.
A series of commands is triggered by the Sun reaching a Solar Zenith Angle (SZA) as prescribed by
the Scientific Authority for the observation site.
Scheduling Hints and Conventions:
•
Schedules are created or altered using the SE command.
•
While in SE, the ^ (shift 6) keys may be used to turn on a flag which causes the Ins and Del
keys to insert or delete full lines.
•
The SA command is a convenient tool for relating the time (GMT) to the SZA.
•
In schedule convention, SZA before noon are considered to be -ve, and after noon they are
+ve.
•
Schedule names follow DOS rules, and automatically are given the SKD extension.
•
An ED is invoked automatically at solar midnight by the SKC routine so an additional ED
need not be programmed into the schedule.
•
Schedules may be “linked”
•
when running a schedule under SKC, the schedule automatically starts the next day.
•
when running a schedule under SK, the schedule does not automatically start the next day.
•
all entries in a schedule are keyed to the SZA, even if they are ‘moon’ measurements.
•
only two character commands can be used in a schedule
•
an integer can be entered at the end of a command line to indicate how many times the
command line is to be repeated i.e. DSZSDS3 will cause the commands DSZSDS to be
repeated three times.
•
the next SZA will interrupt the current command line when the command in process is
finished.
•
the HG sequence is normally inserted such that the internal temperature of the Brewer does
not change more that 2
°
C between HG’s.
HG commands should also be done more frequently in sections of the schedule that have much
micrometer movement i.e. UX and UM.
8. BREWER SCHEDULES – SE, SKC, SK
Содержание BREWER MK IV
Страница 1: ...INSTRUCTION MANUAL OM BA C230 B Mk IV...
Страница 2: ...REVISION HISTORY REV DESCRIPTION DCN DATE APPD B Update 99 16 12 C Update 07 16 10...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIV OPERATOR S MANUAL 22...
Страница 27: ...MKIV OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 57: ...MKIV OPERATOR S MANUAL 54...
Страница 61: ...MKIV OPERATOR S MANUAL 58...
Страница 81: ...MKIV OPERATOR S MANUAL 78...
Страница 91: ...MKIV OPERATOR S MANUAL 88...
Страница 93: ...MKIV OPERATOR S MANUAL 90...
Страница 127: ...MKIV OPERATOR S MANUAL 124...
Страница 133: ...MKIV OPERATOR S MANUAL 130...
Страница 135: ...MKIV OPERATOR S MANUAL 132...