Executing Scripts
When a script is executed, the multiple programs that it contains are executed in the order
specified in the script.
•
Press RUN/PAUSE or STOP to make the PBZ enter the program execution (SEQ
EXECUTION) state.
•
Press RUN/PAUSE to start the execution.
•
The programs in the script are executed according to the instructions in the script.
•
If you press RUN/PAUSE, the execution is paused.
•
If the program is designed to wait for a trigger signal, it enters the WAIT state.
•
When all of the programs in the script complete, the PBZ enters the COMPLETED state.
•
Press ENTER.
The PBZ returns to the state where you can set the DC signal (as if you had pressed the
DC key).
•
To abort execution, press STOP or OUTPUT. If you press OUTPUT, regardless of the
programmed output settings, the output is turned off.
•
If an alarm is generated, the PBZ enters the ALARM state.
TRIG IN: ON
TRIG IN
RUN/PAUSE
ENTER
ENTER
STOP
COMPLETED
RUN
PAUSE
WAIT
ABORT
TRIG IN: ON
TRIG IN
STOP
RUN
PAUSE
WAIT
TRIG IN: ON
TRIG IN
STOP
OUTPUT
OUTPUT
OUTPUT
RUN
PAUSE
WAIT
Start
End
P01
P02
Pnn
ALM CLR
ALM CLR
ALARM
ALARM
ALARM
ALARM
ALARM
ALARM
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
RUN/PAUSE
Содержание PBZ20-20
Страница 11: ...PBZ 11 ...
Страница 18: ...18 PBZ ...
Страница 75: ...Memory Response and Configuration This chapter explains preset memory setup memory response and configura tion ...
Страница 95: ...Sequences This chapter explains sequences ...
Страница 153: ...Maintenance This chapter describes how to clean and inspect the PBZ ...
Страница 156: ...156 PBZ Maintenance ...
Страница 157: ...Appendix A List of Factory Default Settings B Troubleshooting C List of Error Messages D Options ...