150
PBZ
Synchronization Feature
Setting the Signal Source
The following limitations exist when PBZs have been connected and set for synchronization.
DC signal source settings
AC signal source settings
Setting
SYNC MASTER
(master unit)
SYNC SLAVE (slave unit)
SYNC SLAVE END (slave end unit)
DC
Can be set
Can be set
FINE
SOFT
START
Can be set
Can be set
SOFT
STOP
Can be set
Cannot be set (fixed to OFF)
1
1
When you try to set this setting, the error “W005: SYNC SLAVE state” is displayed.
Setting
SYNC MASTER
(master unit)
SYNC SLAVE (slave unit)
SYNC SLAVE END (slave end unit)
AC ON/OFF
Can be set
Cannot be set.
1
(It can only be set to OFF.)
1
When you try to set this setting, the error “W005: SYNC SLAVE state” is displayed.
FUNC
Can be set.
2
2
If you change the slave unit settings when output is on, the phase relationship with the master
unit is cleared. To reset this phase relationship, on the master unit, you have to turn OUTPUT
off and back on again or change the AC ON/OFF setting to OFF and then change it to ON.
AMPL
FREQ
Cannot be set.
1
(It can only be set when output is off.)
PHASE
DUTY
Can be set.
SWEEP
Cannot be set
(Set to OFF)
Cannot be set.
1
(Set to OFF.)
Содержание PBZ20-20
Страница 11: ...PBZ 11 ...
Страница 18: ...18 PBZ ...
Страница 75: ...Memory Response and Configuration This chapter explains preset memory setup memory response and configura tion ...
Страница 95: ...Sequences This chapter explains sequences ...
Страница 153: ...Maintenance This chapter describes how to clean and inspect the PBZ ...
Страница 156: ...156 PBZ Maintenance ...
Страница 157: ...Appendix A List of Factory Default Settings B Troubleshooting C List of Error Messages D Options ...