PBZ
127
Program Examples
4
Se
quences
Editing steps
Editing the program
Item
Step details (“NA” means that the setting
is not edited)
STEP
Step number.
1
2
3
4
5
TIME
Step execution time.
0.1 ms 7 ms
0.1 ms 26 ms
12 ms
DC
DC signal.
0 V
0 V
6 V
15 V
15 V
DC
RAMP
The DC signal ramps up or
down gradually to the DC
setting over time.
OFF
OFF
OFF
ON
OFF
DC
RAMP
START
DC signal start value.
0 V
0 V
0 V
6 V
15 V
DC
RAMP
STOP
DC signal stop value.
0 V
0 V
6 V
15 V
15 V
OUTPU
T
Output on/off state.
ON
ON
ON
ON
ON
TRIG
OUT
Trigger signal output.
OFF
OFF
OFF
OFF
OFF
TRIG IN
Trigger signal input standby.
OFF
OFF
OFF
OFF
OFF
AC
Superimposition of AC signal
on DC signal.
OFF
OFF
OFF
ON
OFF
FUNC
AC signal waveform.
NA
NA
NA
NA
AMPL
AC signal amplitude.
NA
NA
NA
8 Vp-p NA
FREQ
AC signal frequency.
NA
NA
NA
1 kHz
NA
PHASE
AC signal start phase angle.
The resolution is 1 deg.
NA
NA
NA
FREE
NA
DUTY
Square wave duty ratio.
NA
NA
NA
NA
NA
AMPL
SWEEP
AC signal amplitude sweep.
NA
NA
NA
NA
NA
AMPL
START
Start and stop amplitudes for
the AC signal amplitude
sweep.
NA
NA
NA
NA
NA
AMPL
STOP
NA
NA
NA
NA
NA
FREQ
SWEEP
AC signal frequency sweep.
NA
NA
NA
NA
NA
FREQ
MODE
AC signal frequency sweep
mode.
NA
NA
NA
NA
NA
FREQ
START
Start and stop frequencies for
the AC signal frequency
sweep.
NA
NA
NA
NA
NA
FREQ
STOP
NA
NA
NA
NA
NA
Item
PROG
Program number
A number from 1 to 16
TITLE
Program name
SAWTOOTH GEN (ASCII codes from 0x20 to 0x7E)
MODE
CV/CC mode
CV
LOOP
Number of repetitions
(infinity)
Содержание PBZ20-20
Страница 11: ...PBZ 11 ...
Страница 18: ...18 PBZ ...
Страница 75: ...Memory Response and Configuration This chapter explains preset memory setup memory response and configura tion ...
Страница 95: ...Sequences This chapter explains sequences ...
Страница 153: ...Maintenance This chapter describes how to clean and inspect the PBZ ...
Страница 156: ...156 PBZ Maintenance ...
Страница 157: ...Appendix A List of Factory Default Settings B Troubleshooting C List of Error Messages D Options ...