VOLT:LIM:AUTO
Sets whether to enable the voltage limit setting. If enabled, the voltage limit setting is
automatically set to approximately 95 % of the OVP trip point. This cannot be set when the
output is on.
Command
[SOURce:]VOLTage:LIMit:AUTO {ON|OFF|1|0}
[SOURce:]VOLTage:LIMit:AUTO?
Parameter
Value ON(1) Limit the voltage setting
OFF(0) Not limit the voltage setting (default)
For the setting that is applied when *RST is sent, see
.
Response
Returns whether to enable the voltage limit setting in the NR1 form in response to
VOLT:LIM:AUTO?.
Overview
Setup
Overview of Messages
Command
(function search)
Protection and Clear the Alarm
Setting Changes using triggers
Command
(ABC search)
Command
(Sub-system search)
Appendix
Tutorial
PAT-T series Communication Interface Manual
KIKUSUI ELECTRONICS CORP.
90/158