Tutorial
Error Checking
Error/event queue
The SCPI specifications define a standard error reporting scheme, Error/Event Queue. This is
a FIFO (First In First Out) queue, which records errors and events. The maximum number of
errors/events that the PAT-T can record is 255. Each error/event can be read with the
SYSTem:ERRor query.
The response to this query contains a numeric part (error/event number) and a textual
description, such as:
,"Data out of range"
The error/event queue is empty when the *CLS common command is sent, when the last
item in the queue is read, or when the PAT-T is turned on. When the error/event queue is
empty, the query returns the following:
0,"No error"
Displaying communication errors
The PAT-T has a debug trace function. The oldest item among the errors and events (if they
are present) can be displayed on the PAT-T. This function is convenient when you debug
your remote applications.
While an error/event item is displayed on the panel, the normal voltmeter and ammeter are
disabled.v
If the error/event queue is empty, the debug trace function does not display any errors.
Sending the *CLS command clears the communication error display.
If in local mode, the debug trace function is temporarily disabled.
The communication error display can be enabled or disabled with the
SYSTem:CONFigure:TRACe command.
SYSTem:CONFigure:TRACe {ON|OFF}