SYST:CONF:MPR
Sets the state that the PAT-T starts up in when the output state at power-on is set to on.
This cannot be set when the output is on.
This command is available for 8 kW type products only.
Command
SYSTem:CONFigure:MPRiority {CV|CC}
SYSTem:CONFigure:MPRiority?
Parameter
Value CV CV (constant voltage) is prioritized (default)
CC CC (constant current) is prioritized
Response
Returns the state that the PAT-T starts up in when the output state at power-on is set to on
as character data in response to SYST:CONF:MPR?.