SENS:AVER:CLE
It clears all the measurement data (The mean value data: the moving average deviations of
64 times measured in every 4 ms) stored in the measurement data cache, and the
measurement data after 256 ms (New moving average created by another 64 times)
becomes valid.
When the data is acquired while varying the setting of the voltage or the current, transmit
the "SENS:AVER:CLE" command after setting the voltage or the current. The measurement
data becomes valid after 256 ms upon receipt of the command.