PAD-L
Applied Operation
4-17
5.
Turn on the POWER switch.
6.
Adjust the I.
OS
variable resistor on the front panel so that the output
current becomes 0 A when R3 is zero.
C
0
1
2
–
–S
+S
+
3
4
5
6
7
8
9
10
R3
Remove the shorting bar
Control terminals
Output terminals
2-core shielded wire
or twisted-pair wire
Output current Io
≈
Imax · R3
10
[A]
R3
≤
a [
Ω
]
Imax: Maximum output current [A]
(See Capter 6 Table 6-2
"Maximum current value")
a: Constant (depend on model)
Fig, 4-11
Connections for output current control with external resistor
Table 4-2
Constant (depend on model)
Type III
PAD
16-100L
PAD
35-50L
PAD
35-60L
PAD
60-35L
PAD
110-20L
PAD
250-8L
a [
Ω
]
550
1000
1000
1000
1000
1000
Type IV
PAD
35-100L
PAD
60-60L
PAD
110-30L
PAD
250-15L
a [
Ω
]
550
550
550
550
Type V
PAD
35-200L
PAD
60-120L
PAD
110-60L
a [
Ω
]
550
550
550
Содержание PAD-L III Series
Страница 4: ...II PAD L...
Страница 16: ...P 6 Preface PAD L...
Страница 17: ...1 1 PAD L Setup 1 1 Chapter 1 Setup Describes the necessary procedure from unpacking to preparation before use...
Страница 40: ...2 12 Precautions and Preparations for Use PAD L...
Страница 86: ...4 34 Applied Operation PAD L...
Страница 93: ...PAD L Names and Functions of Controls 5 7 22 25 23 26 24 21 Fig 5 6 PAD L series PAD250 15L rear panel...
Страница 95: ...PAD L Names and Functions of Controls 5 9 22 26 23 25 21 24 28 29 Fig 5 8 PAD L series PAD110 60L rear panel...
Страница 98: ...5 12 Names and Functions of Controls PAD L...
Страница 130: ...7 14 Specifications PAD L...