6-8
Maintenance
PAD-L
Voltage system adjustment procedure (continued)
Full scale of the voltmeter
14. Set the output voltage so that the DVM displays the rated output voltage
of the unit.
15. Adjust the V.
FS
variable resistor on the front panel so that the indication
of the unit's voltmeter matches the DVM reading.
Voltage limit
16. Adjust the RV9 variable resistor on the PCB A-200 so that the unit
voltmeter indication is the same when the CURRENT/VOLT. LIMIT
switch is held down and released.
See Figs. 6-5, 6-6, and 6-7.
This completes the voltage system adjustment.
Table 6-1
Maximum voltage value
Type III
PAD
16-100L
PAD
35-50L
PAD
35-60L
PAD
60-35L
PAD
110-20L
PAD
250-8L
Maximum voltage [V]
16.5
35.6
35.6
61
112
260
Type IV
PAD
35-100L
PAD
60-60L
PAD
110-30L
PAD
250-15L
Maximum voltage [V]
35.6
61
115
260
Type V
PAD
35-200L
PAD
60-120L
PAD
110-60L
Maximum voltage [V]
35.6
61
115
PCB A-200
RV9 (Voltage limit ADJ)
RV20 (Maximum variable constant-voltage range ADJ)
Never touch this
variable resistor.
Fig. 6-5
RV9 and RV20 in the TYPE III model
Содержание PAD-L III Series
Страница 4: ...II PAD L...
Страница 16: ...P 6 Preface PAD L...
Страница 17: ...1 1 PAD L Setup 1 1 Chapter 1 Setup Describes the necessary procedure from unpacking to preparation before use...
Страница 40: ...2 12 Precautions and Preparations for Use PAD L...
Страница 86: ...4 34 Applied Operation PAD L...
Страница 93: ...PAD L Names and Functions of Controls 5 7 22 25 23 26 24 21 Fig 5 6 PAD L series PAD250 15L rear panel...
Страница 95: ...PAD L Names and Functions of Controls 5 9 22 26 23 25 21 24 28 29 Fig 5 8 PAD L series PAD110 60L rear panel...
Страница 98: ...5 12 Names and Functions of Controls PAD L...
Страница 130: ...7 14 Specifications PAD L...