4-6
Applied Operation
PAD-L
4.2.2
Output voltage control (II) with external resistor
•
This voltage control is a fail-safe system in which no overshoot occurs when
switching a resistance value.
•
The output voltage Eo and resistance value R2 are inversely proportional, as
shown in Fig. 4-4. If the circuit opens when switching the resistor or due to an
accident, the resistance value become
∞
(infinite) and output becomes zero.
•
The output voltage Eo is determined by R2 and Eref, according to the equation
shown in Fig. 4-5. Eref is set with the VOLTAGE setting knob on the front
panel. (Use a guard cap (supplied) to disable a setting knob on the panel.)
•
The advantage of this applied operation is a fail-safe system in which the output
voltage decreases if the circuit becomes open. The drawback is that it is
impractical when a low voltage is programmed, since a very high resistance
value is required. In real-world applications, 0 to 200 k
Ω
variable resistors are
suitable. (In general, high resistance requires greater attention to temperature
coefficient and noise.)
Eo max
0V
0
Ω
Eo
∞
Resistance value R2
Control (II) with external resistor
Control (I) with e
xternal resistor
Output v
oltage Eo
Set with the knob
on the panel
Fig. 4-4
Control (I) and control (II) with external resistor
Содержание PAD-L III Series
Страница 4: ...II PAD L...
Страница 16: ...P 6 Preface PAD L...
Страница 17: ...1 1 PAD L Setup 1 1 Chapter 1 Setup Describes the necessary procedure from unpacking to preparation before use...
Страница 40: ...2 12 Precautions and Preparations for Use PAD L...
Страница 86: ...4 34 Applied Operation PAD L...
Страница 93: ...PAD L Names and Functions of Controls 5 7 22 25 23 26 24 21 Fig 5 6 PAD L series PAD250 15L rear panel...
Страница 95: ...PAD L Names and Functions of Controls 5 9 22 26 23 25 21 24 28 29 Fig 5 8 PAD L series PAD110 60L rear panel...
Страница 98: ...5 12 Names and Functions of Controls PAD L...
Страница 130: ...7 14 Specifications PAD L...