5-2 Maintenance
KJM6710
5.4
Troubleshooting
The symptoms listed below may not necessarily mean a malfunction.
Check the KJM6710 before requesting repairs.
If the remedy does not solve the problem or if your case does not
match any of the items, contact your Kikusui agent.
Table 5-1
Symptom
Check Items
Description
The KJM6710
does not oper-
ate even when
the power
switch is
turned on.
Is the line voltage
correct?
The allowable line voltage range is
14.5 V to 15.5 V. The line voltage may
drop if the power supply capacity is
insuf
fi
cient.
Is the DC voltage polarity
correct?
For safety reasons, the fuse is
designed to blow when the DC voltage
polarity is reversed. If the fuse is
blown, request for repairs.
Cannot make
measurements.
Is the input amplitude
adequate?
The input impedance must be 50
Ω
,
and the input amplitude must be at
least 0.2 Vp-p. 1-M
Ω
probes cannot
be used.
Is the output impedance
correct?
Check that the output impedance of
the connected device and the cable
impedance are 50
Ω
.
Is the duty cycle of clock
signal outside the speci
fi
-
cations?
The duty cycle speci
fi
cation of the
clock signal is 45:55 to 50:50.
PLL does not
lock.
If the frequency correct?
The PLL of the KJM6710 can lock to
fundamental clock frequencies
between 25 MHz to 30 MHz.
Is the amount of jitter too
large?
The PLL of the KJM6710 can lock up
to 17 %.
Содержание KJM6710
Страница 1: ...KJM6710 Part No Z1 003 052 IA003141 Jan 2004 OPERATION MANUAL Time Interval Jitter Meter ...
Страница 8: ...VI Overvoltage category KJM6710 ...
Страница 12: ...P 2 Preface KJM6710 ...
Страница 52: ...5 4 Maintenance KJM6710 ...
Страница 57: ......
Страница 58: ......
Страница 59: ......
Страница 60: ...TIME INTERVAL JITTER METER KJM6710 OPERATION MANUAL ...