KJM6710
Operation 3-7
3.2
Function Setup
There are two methods for setting the internal functions of the
KJM6710.
1. By using the function setup switches on the rear panel.
2. By transmitting messages via the RS-232C interface.
The table below shows the possible combinations of the setup
through the function setup switches.
*1
Do not turn both RF EDGE switches to
(ON).
After setting the switches, DELAY which is not indicated in the table is
set to AUTO, and SYMMETRY LEVEL (when set to SYMMETRY
MANUAL) is set to 0 V.
The BPS (baud rate) switch setting takes effect the next time the
KJM6710 is powered up. Other switch settings take effect immediately.
All settings except the BPS (baud rate) setting can be con
fi
gured
using RS-232C command messages. The settings that actually take
effect are those speci
fi
ed last.
SW
Operation when set to
(OFF)
Operation when set to
(ON)
BPS (baud rate)
19
200 BPS
9
600 BPS
EQ
OFF
ON
PLL
OFF
ON
TIME CONST
0.1 s
1 s
SYMMETRY
AUTO
MANUAL (0 V)
CLOCK EDGE
positive
negative
RF EDGE
*1
positive
either
negative