RS-232C Control 4-9
PLL
Turns on/off the PLL circuit. Or, queries the current setting.
Program message
• Syntax
Command message: PLL_<{ON|OFF}>
Query message:
PLL?
• Program data
Data format: Characters
Values:
ON:
PLL ON
OFF:
PLL OFF
(Example) Turn on the PLL.
PLL ON
Response message
Returns the current setting in response to PLL? in characters (ON
or OFF).
PLL:STATUS? (PLL:STAT?)
Queries the PLL circuit status.
Program message
• Syntax
Query message: PLL:STATUS?
PLL:STAT?
Response message
Returns the PLL circuit status (LOCK or UNLOCK) in response
to PLL:STAT?.
Message Code
Description
LOCK
The PLL circuit is locked to
the input signal.
UNLOCK
The PLL circuit is not locked
to the input signal.