4-10 RS-232C Control
KJM6710
RF:TRIGGER:EDGE (RF:TRIG:EDG)
Sets the trigger edge polarity of the RF INPUT. Or, queries the
current setting.
Program message
• Syntax
Command message: RF:TRIGGER:EDGE_<{POSITIVE
|NEGATIVE|EITHER}>
RF:TRIG:EDG_<{POS|NEG|EIT}>
Query message:
RF:TRIGGER:EDGE?
RF:TRIG:EDG?
• Program data
Data format: Characters
Values:
POS:
↑
(Rising)
NEG:
↓
(Falling)
EIT:
Rising or falling
(Example) Set the trigger edge polarity to POS.
RF:TRIG:EDG POS
Response message
Returns the current setting in response to RF:TRIG:EDG? in char-
acters (POS, NEG, or EIT).
SYMMETRY:MODE (SYM:MOD)
Sets the symmetry mode. Or, queries the current setting.
Program message
• Syntax
Command message:
SYMMETRY:MODE<{AUTO|OFFSET
|MANUAL}>
SYM:MOD_<{AUT|OFFS|MAN}>
Query message:
SYMMETRY
:MODE?
SYM:MOD?
• Program data
Data format: Characters
Values:
AUT:
Set to AUTO
OFFS: Set to OFFSET
MAN: Set to MANUAL